beautypg.com

Test port input – Atec Agilent-E5063A User Manual

Page 12

background image

12

Test Port Input

Description

Speciication

Typical

Test port input level

Maximum input level

Damage level

+6 dBm

+26 dBm or ± 35 VDC

Crosstalk

100 kHz to 300 kHz

300 kHz to 8.5 MHz

8.5 MHz to 4.34 GHz

4.34 to 6 GHz

6 to 13 GHz

13 to 16 GHz

16 to 18 GHz

–88 dB
–93 dB
–115 dB
–105 dB
–100 dB
–90 dB
–85 dB

Test Port Noise Floor

100 kHz to 8.5 MHz

8.5 to 100 MHz

100 MHz to 4.34 GHz

4.34 to 8.5 GHz

8.5 to 13 GHz

13 to 16 GHz

16 to 18 GHz

–103 dBm/Hz
–126 dBm/Hz
–127 dBm/Hz
–116 dBm/Hz
–115 dBm/Hz
–105 dBm/Hz
–102 dBm/Hz

Compression level
(at maximum test port input level = +6 dBm)

Magnitude

100 kHz to 1 MHz

1 MHz to 4.34 GHz

4.34 to 13 GHz

13 to 18 GHz

± 0.2 dB
± 0.2 dB
± 0.2 dB
± 0.2 dB

Phase

100 kHz to 1 MHz

1 MHz to 4.34 GHz

4.34 to 13 GHz

13 to 18 GHz

± 5 deg.
± 1.5 deg.
± 6 deg.
± 10 deg.