beautypg.com

Dynamic accuracy, Stability – Atec Agilent-E5063A User Manual

Page 14

background image

14

Dynamic accuracy

1

Description

Speciication

Typical

Magnitude

6 dBm

–30 dBm

–100 dBm

–110 dBm

± 0.31 dB
± 0.056 dB
± 3.83 dB

± 5.00 dB

Phase

6 dBm

–30 dBm

–100 dBm

± 11.8 deg
± 0.37 deg
± 33.6 deg

1. Accuracy of the test port input power reading is relative to –10 dBm reference input power level.

Magnitude

Phase

Stability

1

Description

Speciication

Typical

Magnitude
Transmission:

100 kHz to 300 kHz

300 kHz to 6 GHz

6 to 12 GHz

12 to 18 GHz

Relection:

100 kHz to 300 kHz

300 kHz to 6 GHz

6 to 12 GHz

12 to 18 GHz

± 0.02 dB/°C
± 0.01 dB/°C
± 0.025 dB/°C
± 0.04 dB/°C

± 0.02 dB/°C
± 0.02 dB/°C
± 0.035 dB/°C
± 0.05 dB/°C

Phase
Transmission:

100 kHz to 300 kHz

300 kHz to 6 GHz

6 to 12 GHz

12 to 18 GHz

Relection:

100 kHz to 300 kHz

300 kHz to 6 GHz

6 to 12 GHz

12 to 18 GHz

± 0.4 deg/°C
± 0.2 deg/°C
± 0.5 deg/°C
± 0.6 deg/°C

± 0.4 deg/°C
± 0.2 deg/°C
± 0.5 deg/°C
± 0.6 deg/°C

1. Stability is defined as a ratio measurement at the test port.