Port modules, Type-n (50 ohms), 5 mm – Atec Agilent-N4433A User Manual
Page 8: 85 mm
![background image](/manuals/318355/8/background.png)
8
4-port modules
Type-N (50 ohms)
N4431B (RF), Option 020
2
Frequency range
9 kHz
3
to 10 MHz
10 MHz to 1 GHz
1 to 3 GHz
3 to 6 GHz
6 to 8 GHz
8 to 9 GHz
9 to 13.5 GHz
4
Thru paths AB, CD, AD, BC
Directivity (dB)
45
55
52
47
44
42
40
Source match (dB)
36
47
43
42
40
39
31
Reflection tracking (±dB)
0.10
0.03
0.04
0.04
0.05
0.06
0.11
Transmission tracking (±dB)
0.078
0.07
0.10
0.14
0.20
0.22
0.35
Load match (dB)
39
47
45
40
38
35
26
Thru paths AC, BD
Directivity (dB)
45
55
52
47
44
42
40
Source match (dB)
36
47
43
42
40
39
31
Reflection tracking (±dB)
0.10
0.03
0.04
0.04
0.05
0.06
0.11
Transmission tracking (±dB)
0.078
0.07
0.09
0.13
0.15
0.16
0.32
Load match (dB)
39
47
45
40
38
36
28
3.5 mm
N4431B (RF), Option 010
2
Frequency range
9 kHz
3
to 10 MHz
10 MHz to 1 GHz
1 to 3 GHz
3 to 6 GHz
6 to 8 GHz
8 to 9 GHz
9 to 13.5 GHz
4
Thru paths AB, CD, AD, BC
Directivity (dB)
45
57
55
52
50
47
40
Source match (dB)
36
50
47
45
44
43
32
Reflection tracking (±dB)
0.10
0.03
0.03
0.04
0.04
0.05
0.1
Transmission tracking (±dB)
0.078
0.06
0.09
0.12
0.14
0.20
0.33
Load match (dB)
39
47
46
45
44
42
28
Thru paths AC, BD
Directivity (dB)
45
57
55
52
50
47
40
Source match (dB)
36
50
47
45
44
43
32
Reflection tracking (±dB)
0.10
0.03
0.03
0.04
0.04
0.05
0.1
Transmission tracking (±dB)
0.078
0.06
0.08
0.10
0.12
0.14
0.3
Load match (dB)
39
47
46
45
45
43
29
1.85 mm
N4694A (microwave)
1
Frequency range (GHz)
10 to 45 MHz
45 MHz to 2 GHz 2 to 20 GHz
20 to 30 GHz 30 to 40 GHz
40 to 50 GHz 50 to 60 GHz 60 to 67 GHz
Directivity (dB)
33
50
50
46
44
42
41
38
Source match (dB)
25
38
39
35
34
33
30
27
Reflection tracking (±dB)
0.05
0.04
0.04
0.05
0.06
0.07
0.08
0.09
Transmission tracking (±dB)
0.15
0.07
0.1
0.11
0.13
0.15
0.19
0.25
Load match (dB)
25
41
44
42
40
38
36
32
1. When applied power exceeds –5 dBm, calibration results will be degraded from the performance indicated in the table.
2. When applied power exceeds +7 dBm, calibration results will be degraded from the performance indicated in this table.
3. Performance from 9 kHz to 300 kHz is valid only for the E5071C ENA network analyzer with firmware version A.09.10 or higher.
4. 9 to 13.5 GHz range not vaild for the N4431A