beautypg.com

7 mm, 92 mm, 4 mm – Atec Agilent-N4433A User Manual

Page 7

background image

7

7 mm

85091C (RF)

1

Frequency range

300 kHz to 10 MHz

10 MHz to 1 GHz

1 to 3 GHz

3 to 6 GHz

6 to 9 GHz

Directivity (dB)

45

52

56

55

45

Source match (dB)

36

45

44

41

34

Reflection tracking (±dB)

0.1

0.04

0.04

0.07

0.10

Transmission tracking (±dB)

0.078

0.06

0.06

0.13

0.23

Load match (dB)

39

47

47

46

39

N4696B (microwave)

2

Frequency range (GHz) 300 kHz to 2 MHz

2 to 10 MHz

10 to 500 MHz

500 MHz to 2 GHz

2 to 10 GHz

10 to 18 GHz

Directivity (dB)

30

40

46

45

50

42

Source match (dB)

28

35

40

40

42

36

Reflection tracking (±dB)

0.12

0.07

0.05

0.03

0.03

0.05

Transmission tracking (±dB)

0.14

0.11

0.07

0.04

0.07

0.1

Load match (dB)

25

30

36

40

45

39

7-16

85098C (RF)

1

Frequency range

300 kHz to 10 MHz

10 MHz to 1 GHz

1 to 3 GHz

3 to 6 GHz

6 to 7.5 GHz

Directivity (dB)

45

47

50

46

45

Source match (dB)

36

43

43

38

37

Reflection tracking (±dB)

0.1

0.03

0.03

0.05

0.06

Transmission tracking (±dB)

0.078

0.05

0.06

0.08

0.10

Load match (dB)

39

42

43

41

38

2.92 mm

N4692A (microwave)

2

Frequency range (GHz)

10 to 45 MHz, typical

45 MHz to 2 GHz

2 to 20 MHz

20 to 30 GHz

30 to 40 GHz

Directivity (dB)

35

45

43

39

38

Source match (dB)

30

36

35

30

29

Reflection tracking (±dB)

0.10

0.08

0.08

0.10

0.10

Transmission tracking (±dB)

0.10

0.14

0.14

0.20

0.25

Load match (dB)

27

36

37

33

33

2.4 mm

N4693A (microwave)

2

Frequency range (GHz)

10 to 45 MHz, typical

45 MHz to 2 GHz

2 to 10 MHz

10 to 20 GHz

20 to 40 GHz

40 to 50 GHz

Directivity (dB)

32

55

49

45

43

41

Source match (dB)

25

46

42

37

35

30

Reflection tracking (±dB)

0.05

0.03

0.04

0.05

0.06

0.08

Transmission tracking (±dB)

0.10

0.06

0.08

0.11

0.13

0.17

Load match (dB)

24

45

42

40

38

36

1. When applied power exceeds +9 dBm, calibration results will be degraded from the performance indicated in the table.