Input power level, Electrical characteristic corrected performance, Input power limits – Atec Agilent-N4433A User Manual
Page 5: Residual e-terms) 2-port ecal modules
![background image](/manuals/318355/5/background.png)
5
1. If the maximum input power is exceeded when calibrating, compression may occur.
2. When using the PNA-X, the power level can be increased after calibration with minimal impact on measurement accuracy.
3. When mated with male connectors with a 0.77 mm (0.030 in) to 0.86 mm (0.034 in) pin diameter
Input Power Level
Before performing a calibration, make sure the input power and DC levels do not exceed the values indicated in the table below.
Input power limits
ECal module series
Parameter
8509x N4431x
N4432A/N4433A
N469x
Typical maximum input power
1, 2
+9.0 dBm
+7.0 dBm
–7.0 dBm
–5.0 dBm
Typical maximum DC level
±20 Volts
±3 Volts
±3 Volts
±10 Volts
applied to test port
Typical damage level
+20.0 dBm
+20.0 dBm
+20.0 dBm
+10.0 dBm
Operating temperature
The temperature of the ECal module must be within the following temperature range to meet the operating specifications.
• 8509x Series: +20 to +30 °C
• N443xA/B Series: +20 to +30 °C
• N469xA/B Series: +20 to +26 °C
Electrical characteristic corrected performance
(Residual e-terms)
2-port ECal modules
Type-F
85099C (RF)
3
Frequency range
300 kHz to 10 MHz
10 MHz to 300 MHz
300 MHz to 1.3 GHz
1.3 to 3 GHz
Directivity (dB)
45
50
48
43
Source match (dB)
36
48
45
38
Reflection tracking (±dB)
0.1
0.03
0.07
0.15
Transmission tracking (±dB)
0.08
0.08
0.10
0.17
Load match (dB)
39
43
41
39