beautypg.com

Input power level, Electrical characteristic corrected performance, Input power limits – Atec Agilent-N4433A User Manual

Page 5: Residual e-terms) 2-port ecal modules

background image

5

1. If the maximum input power is exceeded when calibrating, compression may occur.
2. When using the PNA-X, the power level can be increased after calibration with minimal impact on measurement accuracy.
3. When mated with male connectors with a 0.77 mm (0.030 in) to 0.86 mm (0.034 in) pin diameter

Input Power Level

Before performing a calibration, make sure the input power and DC levels do not exceed the values indicated in the table below.

Input power limits

ECal module series

Parameter

8509x N4431x

N4432A/N4433A

N469x

Typical maximum input power

1, 2

+9.0 dBm

+7.0 dBm

–7.0 dBm

–5.0 dBm

Typical maximum DC level

±20 Volts

±3 Volts

±3 Volts

±10 Volts

applied to test port
Typical damage level

+20.0 dBm

+20.0 dBm

+20.0 dBm

+10.0 dBm

Operating temperature

The temperature of the ECal module must be within the following temperature range to meet the operating specifications.

• 8509x Series: +20 to +30 °C
• N443xA/B Series: +20 to +30 °C
• N469xA/B Series: +20 to +26 °C

Electrical characteristic corrected performance

(Residual e-terms)

2-port ECal modules

Type-F

85099C (RF)

3

Frequency range

300 kHz to 10 MHz

10 MHz to 300 MHz

300 MHz to 1.3 GHz

1.3 to 3 GHz

Directivity (dB)

45

50

48

43

Source match (dB)

36

48

45

38

Reflection tracking (±dB)

0.1

0.03

0.07

0.15

Transmission tracking (±dB)

0.08

0.08

0.10

0.17

Load match (dB)

39

43

41

39