beautypg.com

I/ix series : product specifications – Atec California-Instruments-i-iX_SeriesII User Manual

Page 6

background image

www.ProgrammablePower.com

174

i/iX Series : Product Specifications

Measurements

Output Parameter

i Series

iX Series

Current Limit Range

Programmable 0 to 100% of range for all ranges

Programmable 0 to 100% of range for all ranges

Current Limit Resolution

0.1 Arms

0.1 Arms

Current Limit Accuracy

± 0.5 A

± 0.5 A

Frequency Range

16.00 - 81.91 Hz (0.01 Hz resolution)

81.0 – 819.1 Hz (0.1 Hz resolution)

820 – 1000 Hz (1 Hz resolution)¹

16.00 - 81.91 Hz (0.01 Hz resolution)

81.0 – 819.1 Hz (0.1 Hz resolution)

820 – 1000 Hz (1 Hz resolution)¹

Frequency Accuracy

± 0.01% of programmed value

± 0.01% of programmed value

DC Offset Voltage

Less than 20 mV with linear load.

Less than 20 mV with linear load.

Output Impedance Range

n/a

Rmin to 1000 mΩ

Lmin to 1000 μH

Output Impedance Resolution

n/a

4 mΩ

4 μH

Output Impedance Accuracy

n/a

± 2% F.S. at 796 μH and 400 mΩ

Output Noise (20 kHz to 1 MHz)

<250 mV rms (typ),

<500 mV rms (max)

<250 mV rms (typ),

<500 mV rms (max)

1 Note: AC voltage in 300V range derates from 300 Vrms max. at 500 Hz to 150 Vrms max. at 1000 Hz.

Measurements : Peak AC Current

Output Parameter

i Series

iX Series

3001i/iX
5001i/iX

110 A for 150 V range,

92 A for 300 V range

110 A for 150V range,

92 A for 300 V range,

10001i/iX

220 A for 150 V range,

184 A for 300 V range

220 A for 150 V range,

184 A for 300 V range

15001i/iX

330 A for 150 V range,

276 A for 300 V range

330 A for 150 V range,

276 A for 300 V range

Crest Factor

Up to 5:1

Up to 5:1

AC Measurements

Parameter

Range

Accuracy (±)

Resolution

Frequency

16.00 - 1000 Hz

2 counts

0.01: 16 to 81.91 Hz

0.1: 82.0 to 819.0 Hz

1: 820 to 1000 Hz

RMS Voltage

0 - 300 Volts

0.25V + 0.1%, <100 Hz

0.25V + 0.2%, 100-1000 Hz

0.01 Volt

RMS Current

0 - 40 Amps

0.25A + 0.1%, <100 Hz

0.25A + 0.2%, 100-1000 Hz

0.001 Amp

Peak Current

0 - 119 Amps

0.5A + 0.2%, <100 Hz

0.5A + 0.5%, 100-1000 Hz

0.01 Amp

VA Power

0 – 6000 VA

10 VA + 0.1%, <100 Hz

20 VA + 0.2%, 100-1000 Hz

1 VA

Real Power

0 – 6000 W

10 W + 0.1%, <100 Hz

20 W + 0.2%, 100-1000 Hz

1 W

Power Factor (>0.2kVA)

0 - 1.00

0.01

Accuracy specifications apply above 100 counts. Current and Power Accuracy specifications are times two for 10001iX and times three for 15001iX. For 10001iX and 15001iX,
resolution decreases by factor of 10, ranges for current and power increases by factor of three. Measurement bandwidth is limited to 16 Khz.

DC Measurements

Parameter

Range

Accuracy (±)

Resolution

Voltage

0 – 400 Volts

0.4 Volts

0.01 Volt

Current

0 – 40 Amps

0.1 Amps

0.001 Amp

Power

0 – 6000 W

20 W

1 W

Accuracy specifications apply above 100 counts. Current and Power Accuracy specifications are times two for 10001iX and times three for 15001iX. For 10001iX and 15001iX,
resolution decreases by factor of 10, ranges for current and power increases by factor of three.