Characteristics, Specifications, Audio generator – Atec Agilent-U8903A User Manual
Page 8: Sine, dual sine, and variable phase

8
Characteristics
Power consumption
250 VA
Power requirements
• 100 to 240 V
ac
• 47 to 63 Hz
Operating environment
• Operating temperature from 0 to 55 °C
• Relative humidity at 20 to 80% RH (noncondensing)
• Altitude up to 3000 m
• Pollution degree 2
• Installation category II
Storage compliance
–55 to 75 °C (23 to 167 °F)
Safety compliance
Certified with:
• IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
• Canada: CAN/CSA-C22.2 No. 61010-1-04
• USA: ANSI/UL 61010-1:2004
EMC compliance
• IEC 61326-1:2005/EN 61326-1:2006
• Canada:
ICES-001:2004
• Australia/New Zealand: AS/NZS CISPR11:2004
Dimensions (W x D x H)
425.6 mm x 405.0 mm x 133.6 mm (16.76 inches x 15.94 inches x 5.25 inches)
Weight
< 8.5 kg (< 18.74 lb) (without cards)
Warranty
• One year for U8903A
• Three months for standard-shipped accessories (see page 13)
Specifications
The following specifications are based on performance with 30 minutes of warm-up time and a temperature from 0 to 55 °C, unless
stated otherwise.
Audio generator
Generated waveform
Sine, dual sine, variable phase, square, noise (Gaussian and rectangular), arbitrary,
DC, multitone, SMPTE IMD (1:1, 4:1, and 10:1), DFD (IEC 60118/IEC 60268)
Sine, dual sine, and variable phase
Frequency
Range
Accuracy
Resolution
5 Hz to 80 kHz
5 ppm
0.1 Hz
Output
Range
(balanced)
Range
(unbalanced/common)
Amplitude
accuracy
Amplitude
resolution
0 V to 16 V
rms
0 V to 8 V
rms
± 1%
1 μV
rms
(limited to five digits of resolution)
Flatness
20 Hz to 20 kHz
5 Hz to 80 kHz
± 0.01 dB
± 0.1 dB
THD + N at 1 kHz, 1 V
rms
,
20 Hz to 20 kHz bandwidth
≤ –95 dB (at 23 °C ± 5°C)
≤ –92 dB (from 0 to 55 °C)
Dual sine ratio range
0 to 100 dB
Phase
–180 ° to 179.99 °
Sweep
Frequency, amplitude, phase