beautypg.com

Atec Agilent-U8903A User Manual

Page 2

background image

2

Table 1. Comparison of frequency range and accuracy

U8903A

HP 8903B

Frequency range

DC and 10 Hz to 100 kHz

20 Hz to 100 kHz

Frequency accuracy

5 ppm (0.0005%)

0.004%

Table 2. Comparison of accuracy and ranges in AC and DC level measurements

U8903A

HP 8903B

AC voltage input range

0 V to 140 V

rms

0.3 mV

rms

to 300 V

rms

AC accuracy

± 1%

± 4%

DC voltage input range

0 to ± 200 V

4 to 300 V

DC accuracy

± 1%

± 1%

Table 3. Comparison of range and residual THD+N measurements

U8903A

HP 8903B

Frequency range

10 Hz to 100 kHz

20 Hz to 100 kHz

Residual THD+N
(signal distortion)
at 80 kHz BW

≤ –101 dB (at 1 kHz, 1 V

rms

),

20 Hz to 20 kHz

–80 dB (or 15 µV),
20 Hz to 20 kHz

Accuracy

± 0.5 dB (< 20 kHz)
± 0.7 dB (< 100 kHz)

± 1 dB (20 Hz to 20 kHz)
± 2 dB (20 to 100 kHz)

Figure 1. The new U8903A audio analyzer (left) offers numerous improvements over the widely used HP 8903B (right).

Replace your 8903B and add
next-generation capabilities

For nearly two decades, the HP 8903B
provided unparalleled versatility and
performance in audio applications.
The U8903A builds on the legacy of
the 8903B by offering faster single-
point measurements (0.4 sec versus
3.0 sec) as well as a wider frequency
range, expanded performance, and
greater functionality (Tables 1, 2, and
3). With the U8903A, you can con-
figure measurements faster through
its graphical user interface (GUI)
and one-button selection of major
operating modes. The color screen
lets you view dual-parameter displays
from one or two channels as well
as graphical displays of sweeps, fre-
quency spectra, and more (Figure 1).

To makes the transition easy, the
next-generation replacement for the
HP 8903B audio analyzer features
a built-in code emulator which
automatically converts 8903B R2D2
code directly into SCPI commands
used by the U8903A. The Agilent
application note Migrating Code from
the 8903B to U8903A
(5990-4135EN)
and the U8903A Programming Guide
(U8903-90027) provide additional
resources to assure you get the most
from this new class of audio analyzer.