Specifications (continued), Audio analyzer (continued) thd + n and sinad, Signal to noise – Atec Agilent-U8903A User Manual
Page 11: Triggering, Amplitude, Frequency

11
Specifications (continued)
Audio analyzer (continued)
THD + N and SINAD
Fundamental frequency range
10 Hz to 100 kHz
Display range
–999.999 to 0 dB
Accuracy
< 20 kHz
< 100 kHz
± 0.5 dB
± 0.7 dB
Input voltage range
< 1 μV to 140 V
rms
Residual distortion
(at 1 kHz, 1 V
rms
, 20 Hz to 20 kHz bandwidth)
≤ –101 dB
3 dB measurement bandwidth
> 130 kHz
Detection
RMS
Display resolution
% up to 3 decimal places (dB up to 2 decimal places)
Signal to noise
Fundamental frequency range
10 Hz to 100 kHz
Display range
–999.999 to 0 dB
Accuracy
< 20 kHz
< 100 kHz
± 0.5 dB
± 0.7 dB
Input voltage range
< 1 μV to 140 V
rms
Residual distortion
(at 1 kHz, 1 V
rms
, 20 Hz to 20 kHz bandwidth)
≤ –101 dB
Triggering
Type
Free run, external
Level
5 V
Minimum trigger high voltage
1.25 V
Maximum trigger low voltage
0.5 V
Input impedance
> 50 kΩ
Amplitude
DC measurement range
0 to ± 200 V
DC accuracy
± 1%
AC accuracy (20 Hz to 100 kHz)
± 1% (at 23 °C ± 5 °C)
± 2% (from 0 °C to 55 °C)
AC level detection
RMS, peak-to-peak, quasi peak
Over range
80% from the peak of the range value will trigger change to next lower range
Frequency
Range
10 Hz to 100 kHz
Minimum input
1 mV (S/N > 40 dB)
Accuracy
5 ppm
Resolution
6 digits