beautypg.com

Specifications (continued), Audio analyzer (continued) thd + n and sinad, Signal to noise – Atec Agilent-U8903A User Manual

Page 11: Triggering, Amplitude, Frequency

background image

11

Specifications (continued)

Audio analyzer (continued)

THD + N and SINAD

Fundamental frequency range

10 Hz to 100 kHz

Display range

–999.999 to 0 dB

Accuracy
< 20 kHz
< 100 kHz

± 0.5 dB
± 0.7 dB

Input voltage range

< 1 μV to 140 V

rms

Residual distortion
(at 1 kHz, 1 V

rms

, 20 Hz to 20 kHz bandwidth)

≤ –101 dB

3 dB measurement bandwidth

> 130 kHz

Detection

RMS

Display resolution

% up to 3 decimal places (dB up to 2 decimal places)

Signal to noise

Fundamental frequency range

10 Hz to 100 kHz

Display range

–999.999 to 0 dB

Accuracy
< 20 kHz
< 100 kHz

± 0.5 dB
± 0.7 dB

Input voltage range

< 1 μV to 140 V

rms

Residual distortion
(at 1 kHz, 1 V

rms

, 20 Hz to 20 kHz bandwidth)

≤ –101 dB

Triggering

Type

Free run, external

Level

5 V

Minimum trigger high voltage

1.25 V

Maximum trigger low voltage

0.5 V

Input impedance

> 50 kΩ

Amplitude

DC measurement range

0 to ± 200 V

DC accuracy

± 1%

AC accuracy (20 Hz to 100 kHz)

± 1% (at 23 °C ± 5 °C)
± 2% (from 0 °C to 55 °C)

AC level detection

RMS, peak-to-peak, quasi peak

Over range

80% from the peak of the range value will trigger change to next lower range

Frequency

Range

10 Hz to 100 kHz

Minimum input

1 mV (S/N > 40 dB)

Accuracy

5 ppm

Resolution

6 digits