beautypg.com

Inputs and outputs – Atec Agilent-N8300A User Manual

Page 6

background image

6

Inputs and
Outputs

Front panel

RF In/Out

Connector

Type-N female, 50 Ω nominal

RF output

Connector

Type-N female, 50 Ω nominal

USB ports

Master (2 ports)

Standard

Compatible with USB 2.0

Connector

USB Type-A female

Output current

0.5 A nominal

Others (not used)

Probe power

Voltage/Current


+15 Vdc, ±7% at 150 mA max nominal

-12.6 Vdc, ±10% at 150 mA max nominal

Headphone jack

Rear panel

SOURCE REF IN

Accepts a 10 MHz reference signal used to
frequency lock the internal timebase

Connector

BNC female, 50 Ω nominal

Input amplitude range

–3.5 to +20 dBm nominal

SOURCE 10 MHz OUT

Outputs the 10 MHz reference signal used
by the internal timebase. Permanently
connected to the ANALYZER 10 MHz IN port

Connector

BNC female, 50 Ω nominal

Output amplitude

+3.9 dBm nominal

Input damage level

+16 dBm

ANALYZER 10 MHz IN

Permanently connected to the
SOURCE 10 MHz OUT port

Connector

BNC female, 50 Ω nominal

Input amplitude range

–5 to +10 dBm nominal

Input frequency

1 to 50 MHz nominal

Frequency range lock

±5 x 10

–6

of specified external reference

input frequency

ANALYZER 10 MHz OUT

Connector

BNC female, 50 Ω nominal

Output amplitude

≥ 0 dBm nominal

Frequency

10 MHz (±10 MHz x frequency reference

accuracy)