Atec Aeroflex-3920 User Manual
Page 17
For the ver y latest specifications visit
w w w.aeroflex.com
FREQUENCY STANDARD I/O
INPUT/OUTPUT CONNECTORS
Impedance
50
Ω
(nominal)
VSWR
Better than 1.2:1 (RF freq. <1.05 GHz)
Better than 1.3:1 (RF freq. >1.05 GHz to
<2.7 GHz)
Input Protection
200 W with warning above 135 W or power
termination temp >100
o
C. Recommended
max of 30 s ON and minimum of 2 min OFF
for power levels above 50 W. (Remove
power immediately when alarm sounds.)
GPIB
Connector Type
24 pin IEEE
Function
IEEE-488.1-1997
ETHERNET
Connector Type
8 Position, RJ-45 100/10 Mbit/s
Function
10/100 Base-T network connection
RS-232
Connector Type
9-pin, D-sub, Male
Baud Rates
300, 600, 1200, 2400, 4800, 9600,
19.2k, 38.4k, 57.6k, 115.2k
Stop Bits
1 or 2
Parity
Odd, even, none
VIDEO
Connector Type
15-pin, D-sub, VGA
Function
VGA for external monitor
IF OUTPUT
Connector Type
BNC
Function
10.7 MHz Receiver IF
Output Level
Proportional to Receive Signal Level
MIC/ACCESSORY
Connector Type
8 position, female DIN
Function
Microphone connection, modulation input,
demod output, PTT operation
PARALLEL PORT
Connector Type
25 position, female D-sub
Function
Printer interface
USB
Connector Type
Twin USB standard connection (rear panel)
Single USB standard connection (front
panel)
Function
USB Version 2.0 interface
TEST PORT
Connector Type
15 position, female 3 tier D-sub
Function
Programmable I/O and voltage output
(optional interface)
AUXILIARY IF INPUT
Connector Type
High-density dual inline
Function
External digital receiver input (optional
interface)
ANT (RF INPUT)
Connector Type
TNC
Function
Receiver input
Impedance
50
Ω
(nominal)
VSWR (with Attenuation
≤10 dB):
Better than 1.44:1 (RF freq.
<1.05 GHz)
Better than 1.58:1 (RF freq.
>1.05 GHz to <2.7 GHz)
Input Protection
10
W with warning above
+17 dBm (Remove power
immediately when alarm
sounds)
GEN (RF INPUT)
Connector Type
TNC
Function
Generator high-level output
Impedance
50
Ω
(nominal)
VSWR (with level <0 dBm):
Better than 1.7:1 (RF freq.
<1.05 GHz)
Better than 1.9:1 (RF freq.
>1.05 GHz to <2.7 GHz)
Input Protection
10 W with warning above
+23 dBm (Remove power
immediately when alarm
sounds)
T/R (RF INPUT/OUTPUT)
Connector Type
Type N
Function
RF power input, generator
low-level output)
INTERNAL FREQUENCY
STANDARD OUTPUT
Frequency
10 MHz (nominal)
Output Level
1 Vpp (nominal) into 50
Ω
Temperature Stability (0 to
50
o
C)
±0.01 ppm
Aging Rate
±0.1 ppm/year after 1 month
continuous use
Warm Up Time
Less than 5 min. to ±0.02 ppm
EXTERNAL FREQUENCY INPUT
Frequency
10 MHz
Input Level
1 to 5 Vpp for sine waves
3.3/5 V TTL for square waves
Connector
BNC socket (10 k
Ω
Input/50
Ω
Output)
LEVEL
Vertical Scales
1, 2, 5, 10, 20 dB per division
Reference Level
0 dB Full Scale (dBr)
Dynamic Range
Greater than 120 dB
Accuracy
±1 dB from 300 Hz to 15 kHz
MARKERS
Number of Markers
2
17