23 Additional technical information about lightwave measurement

error analysis and calibration is available upon request from an
Agilent Technologies representative.

24 Crosstalk effects are included in the dynamic range and

dynamic accuracy specifications.

25 Optical connector repeatability, cable stability, and system drift

are not included. Transmission and transfer function measure-
ments assume a well-matched device that produces no reflection
from its input port.

26 These calibrations are verified with Agilent’s in-house NIST

traceable reference receiver.

27 Tracking accounts for switch repeatability and frequency response

differences between the measurement reference path and test path.

28 For electrical-to-electrical devices:

Return loss (dBe) = –20 log (

ρ)

Transmission (dBe) = 20 log (V2/V1) = 20 log (I2/I1) = 10 log (P2/P1)

29 Lightwave measurement uncertainty is defined as:

Warranted uncertainty = ((system errors)^2 + (random RSS
errors)^2)^0.5. Typical uncertainty is the RSS combination of all
system and random errors.

30 Uncertainty graphs below refer to relative flatness and modulation

bandwidth measurements. An absolute uncertainty value for a
specific data point can be calculated by adding 1.5 dB to the value
found on the uncertainty graphs.

Optical-to-electrical

30

Measurement setup

Calibration type: response & match
Calibration standards:
Lightwave receiver factory calibration data
Agilent 85052D RF calibration kit
Connectors and cables:
HMS-10 lightwave connectors
40 cm single mode fiber
3.5 mm RF connectors
Agilent 85131E RF cable

Measurement uncertainty

29

Transfer function test

Electrical-to-optical

30

Measurement setup

Calibration type: response & match
Calibration standards:
Lightwave source calibration data
Agilent 85052D RF calibration kit
Connectors and cables:
HMS-10 lightwave connectors
40 cm single mode fiber
3.5 mm RF connectors
Agilent 85131E RF cable

Measurement uncertainty

29

Transfer function test

Magnitude
(typical)

Magnitude
(typical)

Phase
(typical)

Phase
(typical)