beautypg.com

Lightwave measurement accuracy summary, Optical transmission and reflection – Atec Agilent-8703A User Manual

Page 8

background image

Lightwave measurement uncertainty is presented in the
following graphs and tables. This covers three types of
measurements: optical (transmission and reflection)
measurements, optical-to-electrical measurements, and
electrical-to-optical measurements. Data is recorded after
an 8703A accuracy enhancement has been performed
using the indicated calibration type. This analysis accounts
for the following errors

23

:

• Residual systematic errors (Table 5)
• System dynamic accuracy (dB from reference)

24

• 3.5 mm connector repeatability

25

• Lightwave source stability
• Lightwave source and receiver factory calibration

uncertainty

26

• Switch repeatability
• Noise

A 10 Hz IF bandwidth, a 16 averaging factor and a
23 ±3°C temperature range are used in all cases. Data
applies to all 8703A internal source configurations of
Table 4.

The following table shows 8703A residual systematic
errors after accuracy enhancement using the same
calibration and setup as stated for each of the three
lightwave measurement types.

Table 5. Residual
lightwave measurement
systematic errors.

Frequency range (GHz)

0.13 to 12.0

12.0 to 20

Optical residual characteristics
(typical)
Lightwave source port return loss

15 dBo

15 dBo

Lightwave receiver port return loss

20 dBo

20 dBo

Transmission tracking

27

±0.55 dB

±0.55 dB

Lightwave directional coupler
test port return loss

37 dBo

37 dBo

Directivity

37 dB

37 dB

Reflection tracking

27

±0.45 dB

±0.45 dB

Electrical residual characteristics
(typical)
Microwave source port return loss

29 dBe

28

29 dBe

Microwave receiver port return loss

30 dBe

30 dBe

8

Lightwave measurement
accuracy summary

Optical transmission and reflection

Measurement setup

Calibration type: response & isolation
Calibration standards:
14.5 dB return loss Fresnel standard
Connectors and cables:
HMS-10 lightwave connectors
40 cm single mode fiber cables

Measurement uncertainty

29

Transmission test

Magnitude

Phase
(typical)

Reflection test

Magnitude
(typical)