Atec Anritsu-S820E User Manual
Page 8
PN: 11410-00749 Rev. A
S820E TDS
Specifications
Site Master™ S820E
Cable and Antenna Analyzer
(continued)
Measurement Accuracy
1
(TOSLK50A-40 or TOSLKF50A-40)
Corrected Measurement Uncertainty
(Transmission from Port 1 to Port 2)
1. Full 2-Port calibration, Default Power, 10 Hz IFBW, No averaging, 10 minute warm-up. TOSLK50A-40 or TOSLKF50A-40 calibration kit.
Load match specification applicable directly at corrected port only. De-rate by approximately 8 dB if using a 3670 series test port cable.
Reflection and Transmission Tracking are typical.
Frequency Range
Directivity
(dB)
Source Match
(dB)
Load Match
(dB)
Reflection
Tracking (dB)
Transmission
Tracking (dB)
1 MHz to 10 GHz
≥ 42
≥ 33
≥ 42
± 0.15
± 0.06
> 10 GHz to 20 GHz
≥ 36
≥ 26
≥ 36
± 0.15
± 0.05
> 20 GHz to 30 GHz
≥ 32
≥ 22
≥ 32
± 0.10
± 0.05
> 30 GHz to 40 GHz
≥ 30
≥ 20
≥ 30
± 0.10
± 0.05
0.1
1
10
-40
-35
-30
-25
-20
-15
-10
-5
0
Return Loss Magnitude Uncertainty (dB)
20 MHz - 8 GHz
8 GHz - 20 GHz
20 GHz - 30 GHz
30 GHz - 40 GHz
1
10
100
-40
-35
-30
-25
-20
-15
-10
-5
0
Return Loss Phase Uncertainty (degrees)
20 MHz - 8 GHz
8 GHz - 20 GHz
20 GHz - 30 GHz
30 GHz - 40 GHz
Return Loss Magnitude (dB)
Return Loss Magnitude (dB)
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
Transmission Magnitude Uncertainty (dB)
20 MHz - 8 GHz
8 GHz - 14 GHz
14 GHz - 40 GHz
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
Transmission Phase Uncertainty (degrees)
20 MHz - 8 GHz
8 GHz - 14 GHz
14 GHz - 40 GHz
Transmission Magnitude (dB)