Atec Anritsu-S820E User Manual
Page 6

PN: 11410-00749 Rev. A
S820E TDS
Specifications
Site Master™ S820E
Cable and Antenna Analyzer
(continued)
Measurement Accuracy
1
(OSLN50A-18 or OSLNF50A-18, TOSLN50A-18 or TOSLNF50A-18)
Corrected Measurement Uncertainty
(Transmission from Port 1 to Port 2)
1. Full 2-Port calibration, Default Power, 10 Hz IFBW, No averaging, 10 minute warm-up. OSLN50A-18, OSLNF50A-18, TOSLN50A-18, or TOSLNF50A-18 calibration kit.
Load match specification applicable directly at corrected port only. De-rate by approximately 8 dB if using a 3670 series test port cable.
Reflection and Transmission Tracking are typical.
Frequency Range
Directivity
(dB)
Source Match
(dB)
Load Match
(dB)
Reflection
Tracking (dB)
Transmission
Tracking (dB)
1 MHz to 6 GHz
≥ 42
≥ 33
≥ 42
± 0.15
± 0.06
> 6 GHz to 9 GHz
≥ 37
≥ 33
≥ 37
± 0.15
± 0.06
> 9 GHz to 18 GHz
≥ 33
≥ 26
≥ 33
± 0.15
± 0.06
0.1
1
10
-40
-35
-30
-25
-20
-15
-10
-5
0
Return Loss Magnitude Uncertainty (dB)
20 MHz - 6 GHz
6 GHz - 8 GHz
8 GHz - 18 GHz
1
10
100
-40
-35
-30
-25
-20
-15
-10
-5
0
Return Loss Phase Uncertainty (degrees)
20 MHz - 6 GHz
6 GHz - 8 GHz
8 GHz - 18 GHz
Return Loss Magnitude (dB)
Return Loss Magnitude (dB)
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
Transmission Magnitude Uncertainty (dB)
20 MHz - 8 GHz
8 GHz - 14 GHz
14 GHz - 18 GHz
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
Transmission Phase Uncertainty (degrees)
20 MHz - 8 GHz
8 GHz - 14 GHz
14 GHz - 18 GHz
Transmission Magnitude (dB)
Transmission Magnitude (dB)