beautypg.com

Lmr master™ s412e specifications, Power meter, Cw signal generator – Atec Anritsu-S412E User Manual

Page 9

background image

Page 10 of 28

LMR Master™ S412E Specifications

Power Meter

Frequency

Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #, Full Band

Amplitude

Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale

Average

Acquisition Fast/Med/Slow, # of Running Averages

Limits

Limit On/Off, Limit Upper/Lower

Frequency Range

10 MHz to 1.6 GHz (Standard), 10 MHz to 6 GHz (Option 6)

Span

1 kHz to 100 MHz

Display Range

-140 dBm to +30 dBm, ≤ 40 dB span

Measurement Range

-120 dBm to +26 dBm

Offset Range

0 dB to +100 dB

VSWR

2:1 typical

Maximum Power

+35 dBm without attenuator

Accuracy

Same as Spectrum Analyzer

Application Options

Impedance (50 Ω, 75 Ω, Other)

High Accuracy Power Meter (Option 0019) (Requires external USB Power Sensor(s))

Amplitude

Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale

Average

# of Running Averages, Max Hold

Zero/Cal

Zero On/Off, Cal Factor (Center Frequency, Signal Standard)

Limits

Limit On/Off, Limit Upper/Lower

Power Sensor Model

PSN50

MA24104A/05A

MA24106A

MA24108/18/26A

Description

High Accuracy

RF Power Sensor

Inline High/Peak

Power Sensor

High Accuracy

RF Power Sensor

Microwave

USB Power Sensor

Frequency Range

50 MHz to 6 GHz

600 MHz to 4 GHz

(MA24104A)

350 MHz to 4 GHz

(MA24105A)

50 MHz to 6 GHz

10 MHz to 8 GHz

(MA24108A)

10 MHz to 18 GHz

(MA24118A)

10 MHz to 26 GHz

(MA24126A)

Connector

Type N(m), 50 Ω

Type N(m), 50 Ω

(MA24104A)

Type N(f), 50 Ω

(MA24105A)

Type N(m), 50 Ω

Type N(m), 50 Ω

(MA24108/18A)

Type K(m), 50 Ω

(MA24126A)

Dynamic Range

-30 dBm to +20 dBm

(.001 mW to 100 mW)

+3 dBm to +51.76 dBm

(2 mW to 150 W)

-40 dBm to +23 dBm

(0.1 µW to 200 mW)

-40 dBm to +20 dBm

(0.1 µW to 100 mW)

VBW

100 Hz

100 Hz

100 Hz

50 kHz

Measurand

True-RMS

True-RMS

True-RMS

True-RMS, Slot Power,

Burst Average Power

Measurement Uncertainty

±0.16 dB

1

±0.17 dB

2

±0.16 dB

1

±0.18 dB

3

Datasheet

(for complete specifications)

11410-00414

11410-00483

(MA24104A)

11410-00621

(MA24105A)

11410-00424

11410-00504

Notes:

1) Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than -20 dBm

with zero mismatch errors.

2) Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load.

Measurement results referenced to the input side of the sensor.

3) Expanded uncertainty with K=2 for power measurements of a CW signal greater than -20 dBm with zero mismatch errors.

CW Signal Generator

Setup Parameters

Generator

On/Off

Tx Output Level

-

120 dBm to 0 dBm

Tx Pattern

CW

RF Characteristics

Power Level Accuracy

2.0 dB (CW Pattern, temperature range 15 °C to 35 °C, -120 to 0 dBm) Typical

Frequency Range

500 kHz to 1.6 GHz

Frequency Accuracy

Same as Spectrum Analyzer