Conformance with 3GPP TS 25.141 base station requirements for a manufacturing environment
Sub-clause
Name
3GPP required test
Instrument tolerance
Supplemental
instrument tolerance
interval
information
(as of June 2002)
6.2.1
Maximum output power
±0.7 dB (95%)
±0.29 dB (95%)
±0.63 dB (100%)
6.2.2
CPICH power accuracy
±0.8 dB (95%)
±0.30 dB (95%)
–10 dB CDP
6.3.4
Frequency error
±12 Hz (95%)
±10 Hz (100%)
Freq ref locked
6.4.2
Power control steps
1-dB step
±0.1 dB (95%)
±0.03 dB (95%)
Test Model 2
0.5-dB step
±0.1 dB (95%)
±0.03 dB (95%)
Test Model 2
Ten 1-dB steps
±0.1 dB (95%)
±0.03 dB (95%)
Test Model 2
Ten 0.5-dB steps
±0.1 dB (95%)
±0.03 dB (95%)
Test Model 2
6.4.3
Power dynamic range
±1.1 dB (95%)
±0.50 dB (95%)
6.4.4
Total power dynamic range
±0.3 dB (95%)
±0.015 dB (95%)
Ref –35 dBm
at mixer
6.5.1
Occupied bandwidth
±100 kHz (95%)
±38 kHz (95%)
10 averages
6.5.2.1
Spectrum emission mask
±1.5 dB (95%)
±0.59 dB (95%)
Absolute peak
6.5.2.2
ACLR
5 MHz offset
±0.8 dB (95%)
±0.34 dB (95%)
±0.93 dB (100%)
10 MHz offset
±0.8 dB (95%)
±0.40 dB (95%)
±0.82dB (100%)
6.7.1
EVM
±2.5% (95%)
±1.0% (95%)
Range 15 to 20%
6.7.2
Peak code domain error
±1.0 dB (95%)
±1.0 dB (nominal)
Conditions
25 to 35 °C
Derived tolerances
95th percentile
100% limit tested
Calibration uncertainties included