Atec Agilent-E4406A User Manual
Page 18
18
Electrostatic
In accordance with IEC 801-2/1991, an discharge air
discharge of up to 8 kV, or a contact discharge of up to
4 kV, will not cause any change of instrument state or
measurement data. However, discharges to center pins of
front or rear panel connectors might cause damage to the
associated circuitry.
Power requirements
Voltage, frequency
90 to 132 V rms, 47 to 440 Hz
195 to 250 V rms, 47 to 66 Hz
Power consumption, ON
< 350 W
Power consumption,
< 20 W
standby
Weight
Net
19 kg (42 lb) (nominal)
20 kg (44 lb) with baseband
IQ inputs
Shipping
39 kg (86 lb) (nominal)
Dimensions
177 mm H x 426 mm W x
432 mm D
(7.0 in H x 16.8 in W x 17 in D)
Front panel
RF input
Connector
Type N female
Impedance
50
Ω
(nominal)
VSWR
20 to 2205 MHz
≤
1.4:1
(
≤
1.24:1, typical)
2205 MHz to 4 GHz
≤
1.6:1
(
≤
1.4:1, typical)
50 MHz
≤
1.4:1
(
≤
1.08:1, typical)
Baseband I/Q inputs
Connectors
(4 each I, Q, I
–
, Q
–
) BNC female
Balanced input
600
Ω
, 1 M
Ω
(nominal)
impedance
(switchable)
(4 connectors:
I, Q, I
–
, and Q
–
)
Unbalanced input
50
Ω
, 1 M
Ω
(nominal)
impedance
(switchable)
(2 connectors: I and Q)
VSWR
≤
1.4:1 (
≤
1.08:1, typical)
50
Ω
impedance only
Probe pwr
Voltage/current
+15 Vdc, ±7% at 150 mA
maximum
–12.6 Vdc, ±10% at 150 mA
maximum
Rear panel
10 MHz OUT
Connector
BNC female
Impedance
50
Ω
(nominal)
Output amplitude
≥
0 dBm (nominal)
EXT REF IN
Connector
BNC female
Impedance
50
Ω
(nominal)
Input amplitude range
–5 to +10 dBm (nominal)
Maximum DC level
±28 Vdc
Frequency
1 MHz to 30 MHz, selectable
Frequency lock range
±5 x 10–6 of the specified
external reference input
frequency
TRIGGER IN
Connector
BNC female
Impedance
–10 k
Ω
(nominal)
Trigger level
–5 V to +5 V
TRIGGER 1 OUT and TRIGGER 2 OUT
Connector
BNC female
Impedance
50 k
Ω
(nominal)
Trigger level
0 V to +5 V (no load)
MONITOR output
Connector
VGA compatible, 15-pin mini
D-SUB
Format
VGA (31.5 kHz horizontal,
60 Hz vertical sync rates,
noninterlaced)
Resolution
640 x 480
PARALLEL interface
Allows printing to compatible printers
GPIB interface
Allows communication with compatible devices
Note: Instrument noise sidebands and spurious responses might be
affected by the quality of the external reference used.