Power vs. time and power control measurement

Absolute power measurement

Using 5 MHz resolution bandwidth

Accuracy

0 to –20 dBm

±0.7 dB (nominal)

–20 to –60 dBm

±1.0 dB (nominal)

Relative power measurement

Accuracy

Step range ± 1.5 dB ±0.1 dB (nominal)

Step range ± 3.0 dB ±0.15 dB (nominal)

Step range ± 4.5 dB ±0.2 dB (nominal)

Step range ± 26.0 dB ±0.3 dB (nominal)

Multicarrier power measurement

This measurement is used for adjusting multicarrier power
amplifiers to transmit well balanced multiple carriers. The
measurement is similar to a combination of those for ACPR
and intermodulation distortion product measurements
giving in-channel and out-of-channel performance results.
The results are displayed for the different frequency offsets
either in relative power to the carrier in dBc or in absolute
power in dBm.

Minimum carrier power

–15 dBm (nominal)

at RF input

ACPR dynamic range,

RRC weighted, 3.84 MHz

two carriers

noise bandwidth

5 MHz offset

–64 dB (nominal)

10 MHz offset

–68 dB (nominal)

ACPR accuracy, two carriers

5 MHz offset,

±0.70 dB (nominal)

–48 dBc ACPR

Spectrum emission mask measurement

The spectrum emission mask measurement measures
the in-channel and out-of-channel spurious emissions to
provide useful figures of merit for spectral regrowth and
emissions produced by components and circuit blocks. Up
to five pairs of offsets/regions can be defined in which the
user can specify the start and stop frequencies, resolution
bandwidth, and the start and stop amplitudes of the mask.

Minimum power

–20 dBm (nominal)

at RF input

Dynamic range, relative

2.515 MHz offset

–77.9 dB (–82.8 dB, typical)

1980 MHz region

–72.2 dB (–77.2 dB, typical)

Sensitivity, absolute

2.515 MHz offset

–88.9 dBm (–93.9 dBm, typical)

1980 MHz region

–72.9 dBm (–77.9 dBm, typical)

Accuracy

Display = Abs Peak Pwr ±0.60 dB (±0.40 dB, typical)

Display = Rel Peak Pwg ±0.25 dB

Occupied bandwidth measurement

Occupied bandwidth (OBW) measurement measures the
frequency bandwidth corresponding to 99 percent of the
total transmitted power.

Minimum carrier power

–20 dBm (nominal)

at RF input

Frequency resolution

100 Hz

Frequency accuracy

(nominal)

8

1.4%

N

avg