Frequency – Atec Agilent-E4406A User Manual
Page 2

2
Frequency
Frequency range
RF input
7 to 314 MHz and 329 MHz
to 4 GHz
Baseband IQ inputs
0 Hz to 5 MHz
Frequency spans
Baseband IQ inputs
5 Hz to 5 MHz
(Baseband I or Q inputs)
10 Hz to 10 MHz
(Composite I/Q)
Frequency setting resolution
1 Hz
Frequency reference
Accuracy
±[(time since last adjustment x
aging rate) + temperature
stability + calibration accuracy]
Initial calibration accuracy ±5 x 10
-8
Settability
±2 x 10
-9
Aging rate
During any 24 hrs
±5 x 10
-10
(nominal)
following 24-hr warm-up
Per year
±1 x 10
-7
(nominal)
Temperature stability
±5 x 10
-8
variation from
frequency at +25 °C over the
temperature range of 0 to +55 °C
Warm-up time
1 hour (nominal)
Residual responses
RF input
50
Ω
input terminated, 0 dB input attenuation,
+18 dB ADC gain
20 MHz to 2 GHz
≤
–85 dBm
2 GHz to 4 GHz
≤
–80 dBm
Baseband IQ inputs
50
Ω
input terminated
0 to 5 MHz
≤
–90 dBm
Noise Sidebands (RF Input)
673.6 MHz
Offset
Specifications
Supplemental
100 Hz
≤
–85 dBc/Hz
1 kHz
≤
–92 dBc/Hz
10 kHz
≤
–102 dBc/Hz
100 kHz
≤
–131 dBc/Hz
600 kHz
≤
–138 dBc/Hz
1.2 MHz
≤
–141 dBc/Hz
6.0 MHz
≤
–145 dBc/Hz
10.0 MHz
≤
–145 dBc/Hz
960 MHz
Offset
Specifications
Supplemental
100 Hz
≤
–81 dBc/Hz
1 kHz
≤
–87 dBc/Hz
10 kHz
≤
–96 dBc/Hz
100 kHz
≤
–125 dBc/Hz
600 kHz
≤
–136 dBc/Hz
1.2 MHz
≤
–140 dBc/Hz
6.0 MHz
≤
–146 dBc/Hz
10.0 MHz
≤
–146 dBc/Hz
1990 MHz
Offset
Specifications
Supplemental
100 Hz
≤
–75 dBc/Hz
1 kHz
≤
–82 dBc/Hz
10 kHz
≤
–86 dBc/Hz
100 kHz
≤
–118 dBc/Hz
600 kHz
≤
–132 dBc/Hz
1.2 MHz
≤
–137 dBc/Hz
6.0 MHz
≤
–141 dBc/Hz
10.0 MHz
≤
–141 dBc/Hz
Noise Sidebands
1
(Baseband IQ Inputs)
0 to 5 MHz
Offset
Specifications
Supplemental
1 kHz
≤
–120 dBc/Hz
10 kHz
≤
–133 dBc/Hz
100 kHz
≤
–134 dBc/Hz
1.0 MHz
≤
–135 dBc/Hz (nominal)
5.0 MHz
≤
–135 dBc/Hz (nominal)
1.
No DC offset applied