Measurements – Atec Agilent-E4406A User Manual
Page 4
Amplitude linearity
Baseband IQ inputs
0 to –35 dB below range
±0.17 dB
–35 to –55 dB below range ±1.0 dB
Displayed average noise level
RF input
Input terminated in 50
Ω
, 0 dB attenuation, 1 kHz RBW,
10 kHz span, +18 dB ADC gain
7 to 20 MHz
–103 dBm (–111 dBm, typical)
20 to 2000 MHz
–106 dBm (–111 dBm, typical)
2000 to 2700 MHz
–103 dBm (–108 dBm, typical)
2700 to 4000 MHz
–98 dBm (–104 dBm, typical)
Baseband IQ inputs
Input terminated in 50
Ω
, 1 kHz RBW, 1 kHz to 5 MHz
+13 dBm range
–95 dBm (–100 dBm, typical)
+7 dBm range
–
(–105 dBm, typical)
+1 dBm range
–
(–108 dBm, typical)
–5 dBm range
–106 dBm (–110 dBm, typical)
DC offset
Baseband IQ inputs
After auto-zero
< –40 dB below range
(–55 dB below range, typical)
Compensation for
≤
±2.0 Vdc (offset accuracy
customer
±2.0% of range (nominal))
DC offset
Channel match
Baseband IQ inputs
Amplitude match
±0.25 dB
0 to 5.0 MHz
Phase match
±2.0 degrees
0 to 5.0 MHz
Crosstalk
Baseband IQ inputs
Input impedance = 50
Ω
< –60 dB
Input impedance = 600
Ω
< –52 dB
Common mode rejection
Baseband IQ inputs
600
Ω
balanced inputs
0 to 0.5 MHz
< –50 dB
> 0.5 to 5.0 MHz
< –35 dB
Measurements
Waveform measurement
Range at RF input
Maximum
+30 dBm (1 W)
Minimum
Displayed average noise level
Range at IQ input
Maximum (50
Ω
input) +13 dBm (20 mW)
Maximum 1
V
(600
Ω
, 1 M
Ω
input)
Minimum
Displayed average noise level
Sweep time range
RBW < 7.5 MHz
10 µs to 200 ms
RBW < 1 MHz
10 µs to 400 ms
RBW < 100 kHz
10 µs to 2 s
RBW < 10 kHz
10 µs to 20 s
Time record length
2 to > 900,000 points (nominal)
Resolution bandwidth
1, 1.5, 2, 3, 5, 7.5, 10 sequence,
or arbitrary bandwidth (user-definable)
Gaussian filter
10 Hz to 8 MHz
Flat filter
10 Hz to 10 MHz
Averaging
Average number
1 to 10,000
Average mode
Exponential, repeat
Average type
Power average (RMS),
log-power average (video),
maximum, minimum
Displays
RF input
Signal envelope, I/Q waveform,
I/Q polar
Baseband IQ input
Signal envelope, linear envelope,
I/Q waveform, I and Q
waveform, I/Q polar
Markers
Normal, delta, band power
Nominal dynamic range
(800 MHz to 2 GHz)
–50
–60
–70
–80
–90
–100
–110
–120
–130
–80
–70
–60
–50
–40
–30
–20
–10
0
Mixer level (dBm)
Dynamic range (dB)
Sensitivity (1 Hz RBw)
TOI
4