2 Chest Compression System – Instructions for Use

100666-00 F Valid from CO J2547, © 2011 JOLIFE AB

31

9.7 Electromagnetic environmental declaration

Guidance and manufacturer's declaration - electromagnetic emissions

LUCAS 2 is intended for use in the electromagnetic environment specified below. The customer or the operator of
LUCAS 2 must make sure that it is used in the correct environment.

Emissions test

Compliance

Electromagnetic
environment - guidance

RF emissions CISPR 11

Group 1

LUCAS 2 uses RF energy only for its
internal operation. This makes its RF
emissions very low and not likely to
cause interference with other electronic
equipment near LUCAS 2.

RF emissions CISPR 11

Class B

LUCAS 2 is suitable for use in all
buildings including domestic homes
and places directly connected to the
public low-voltage Power Supply
network that supplies buildings used
for domestic purposes.

Harmonic emissions IEC 61000-3-2

Class A

Voltage fluctuations/ flicker emissions
IEC 61000-3-3

Complies

Guidance and manufacturer's declaration - electromagnetic immunity

LUCAS 2 is intended for use in the electromagnetic environment specified below. The customer or the operator of
LUCAS 2 must make sure that it is used in the correct environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic
environment - guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

+/- 6 kV contact
+/- 8 kV air

+/- 6 kV contact
+/- 8 kV air

Floors must be wood, concrete or
ceramic tile. If there is synthetic
material on the floor, the relative
humidity must be 30% or more.

Electrical fast
transient / Burst
IEC 61000-4-4

+/- 2 kV for Power
Supply lines
+/- 1 kV for input/output lines

+/- 2 kV for Power
Supply lines
n/a. for input/output lines

The mains power quality must be
that of a typical commercial or
hospital environment.

Surge
IEC 61000-4-5

+/- 1 kV differential mode
+/- 2 kV common mode

+/- 1 kV differential mode
n/a. for common mode

The mains power quality must be
that of a typical commercial or
hospital environment.

Voltage dips, short in-
terruptions and voltage
variations on Power
Supply input lines
IEC 61000-4-11

<5 % UT (>95 % dip in UT)
for 0.5 cycle

40 % UT (60 % dip in UT)
for 5 cycles

70 % UT (30 % dip in UT)
for 25 cycles

<5 % UT (>95 % dip in UT))
for 5 sec

<5 % UT (>95 % dip in UT)
for 0.5 cycle

40 % UT (60 % dip in UT)
for 5 cycles

70 % UT (30 % dip in UT)
for 25 cycles

<5 % UT (>95 % dip in UT))
for 5 sec

The mains power quality must be
that of a typical commercial or
hospital environment. If the user of
the [Equipment or System] requires
continued operation during power
mains interruptions, JOLIFE recom-
mends that the [Equipment or
System] is energized from a Power
Supply or Battery that cannot be
interrupted.

Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8

30 A/m

30 A/m

The power frequency magnetic
fields must be at levels that are
characteristic of a typical location in
a typical commercial or hospital
environment

NOTE UT is the a.c. mains voltage prior to application of the test level.

Guidance and manufacturer's declaration - electromagnetic immunity

LUCAS 2 is intended for use in the electromagnetic environment specified below. The customer or the operator of LUCAS
2 must make sure that it is used in the correct environment.