LUCAS
TM
2 Chest Compression System – Instructions for Use
100666-00 F Valid from CO J2547, © 2011 JOLIFE AB
11
3.6 General safety precautions
Caution - use only approved 
accessories
Use only JOLIFE AB-approved accessories 
with LUCAS. LUCAS may not operate 
correctly if you use accessories that are not 
approved. Use only LUCAS Batteries and 
the LUCAS Power Supply that are designed 
for LUCAS. If you use other batteries or 
power supply you can cause permanent 
damage to LUCAS. This also voids the 
warranty.
Caution - liquid
Do not immerse LUCAS in liquid. The device 
can be damaged if liquid enters the hood. 
3.7 Battery
WARNING - LOW BATTERY
When the orange Battery LED shows an 
intermittent light, do one of these:
• Replace the Battery with one that is
charged.
• Connect the external LUCAS Power
Supply.
Caution - keep Battery installed
The Battery must always be installed for 
LUCAS to be able to operate, also when 
powered by the external Power Supply.
To minimize interruptions, we recommend to 
always have a charged spare LUCAS Battery 
in the Carrying Bag.
3.8 Operation
WARNING - UNSATISFACTORY 
POSITION 
Start manual CPR again if it is not possible to 
position LUCAS safely and correctly on the 
patient's chest.
WARNING - INCORRECT POSITION 
OVER CHEST
If the pressure pad is not in the correct 
position in relation to the sternum, there is an 
increased risk of damage to the rib cage and 
the internal organs. Also, the patient's blood 
circulation is compromised.
WARNING - INCORRECT START 
POSITION
The patient's blood circulation is 
compromised if the pressure pad presses 
down too heavily or too lightly on the chest. 
Push the ADJUST key and adjust the height 
of the Suction Cup immediately.
WARNING - CHANGED POSITION 
DURING OPERATION
If the position of the Suction Cup changes
during operation or during defibrillation, 
immediately push ADJUST and adjust the 
position. Always use the LUCAS Stabilization 
Strap to help secure the correct position. 
Caution - defibrillation electrodes
Position the defibrillator electrodes and wires 
so that they are not under the Suction Cup. If 
there are already electrodes on the patient, 
make sure that they are not under the 
Suction Cup. If they are, you must apply new 
electrodes.
Caution - gel on chest
If there is gel on the patient's chest (e.g. from 
ultrasound examination), the position of the 
Suction Cup can change during use.Remove 
all gel before you apply the Suction Cup.
Caution - Stabilization Strap application
Delay the application of the LUCAS 
Stabilization Strap if this prevents or delays 
any medical treatment of the patient.
Caution - adjunctive therapies
The use of other medical equipment or drugs 
in conjunction with LUCAS can affect the 
treatment. Always consult the Instructions 
for Use for the other equipment and/or drugs 
to make sure that they are appropriate for 
use in conjunction with CPR.
WARNING - ECG interference
Chest compressions interfere with ECG 
analysis. Push PAUSE before you start the 
ECG analysis. Make the interruption as short 
as possible. Push ACTIVE (continuous) or 
ACTIVE (30:2) to start the compressions 
again.