Physio-Control LUCAS 2 User Manual
Page 16
LUCAS
TM
2 Chest Compression System – Instructions for Use
100666-00 F Valid from CO J2547, © 2011 JOLIFE AB
16
5.3 Apply to patient
1. Remove the LUCAS Back Plate from the
Carrying Bag.
2. Stop manual CPR.
3. Make sure that you support the patient's
head.
4. Carefully put the LUCAS Back Plate
under the patient, immediately below the
arm pits. Use one of these procedures:
a. Hold the patient's shoulder and lift the
patient's upper body a small distance,
b. Roll the patient from side to side.
Note: An accurate position of the
Back Plate makes it easier and faster
to position the Suction Cup correctly.
5. Start manual CPR again.
6. Hold the handles on the support legs to
remove the LUCAS Upper Part from the
bag. Pull the release rings once to make
sure that the claw locks are open.
7. Let go of the release rings.