Physio-Control LUCAS 2 User Manual
Page 24
LUCAS
TM
2 Chest Compression System – Instructions for Use
100666-00 F Valid from CO J2547, © 2011 JOLIFE AB
24
6.1 Cleaning routines
Clean all surfaces and straps with a soft cloth
and warm water with a mild cleaning agent
or disinfectant agent, e.g.
• 70% isopropyl alcohol solution
• 45% isopropyl alcohol with added
detergent
• Quaternary ammonium compound
• 10% bleach
Follow the handling instructions from the
manufacturer of the disinfectant.
Caution - liquid
Do not immerse LUCAS in liquid. The device
can be damaged if liquid enters the hood.
Allow LUCAS to dry before you pack it into
the bag.
6.2 Remove and install the
Suction Cup
• Pull the Suction Cup off the black mount-
ing tube.
• Discard the Suction Cup as contaminated
medical waste.
• Bend a new Suction Cup onto the black
mounting tube.
• Make sure the Suction Cup is safely
attached on the mounting tube.
6.3 Remove and attach the
Patient Straps
Remove:
1. Open the Patient Straps and pull them
out from the metal rings on the LUCAS
support legs.
Clean according to 6.1
.
Install:
1. Thread the Patient Straps through the
metal holder on the LUCAS support legs.
2. Fold the Patient Strap so that the symbol
is visible.
3. Press the strap parts firmly together.