Physio-Control LUCAS 2 User Manual
Page 18

LUCAS
TM
2 Chest Compression System – Instructions for Use
100666-00 F Valid from CO J2547, © 2011 JOLIFE AB
18
1. Use your finger to make sure that the
lower edge of the Suction Cup is imme-
diately above the end of the sternum
If necessary, move the device by pulling 
the support legs to adjust the position.
2. Adjust the height of the Suction Cup to
set the Start Position.
a. Make sure that LUCAS is in the 
ADJUST mode.
b. Push the Suction Cup down with two
fingers until the pressure pad touches 
the patient's chest without 
compressing the chest.
c. Push PAUSE to lock the Start
Position - then remove your fingers 
from the Suction Cup. 
d. Check for proper position. If not, push
ADJUST, pull up the Suction Cup to 
readjust the central and/or height 
position for a new Start Position. 
Push PAUSE.
e. Push ACTIVE (continuous) OR
ACTIVE (30:2) to start the 
compressions.
Note: If the pressure pad is pushed down 
too hard, or too loose to the chest, LUCAS 
will adjust the pressure pad to the correct 
Start Position (within a range of 30 mm / 1.2 
inches).
WARNING - UNSATISFACTORY 
POSITION 
Start manual CPR again if it is not possible to 
position LUCAS safely and correctly on the 
patient's chest.
WARNING - TOO SMALL PATIENT
If you cannot enter the PAUSE mode or 
ACTIVE mode when the pressure pad 
touches the patient's chest and LUCAS 
alarms with 3 fast signals. Start manual 
compressions again.