4 first use preparations – Physio-Control LUCAS 2 User Manual
Page 12

LUCAS
TM
2 Chest Compression System – Instructions for Use
100666-00 F Valid from CO J2547, © 2011 JOLIFE AB
12
WARNING - PATIENT INJURY
Do not let the patient or the device stay 
unattended when LUCAS operates.
Caution - keep your fingers away
Do not put your hands on or below the 
Suction Cup when LUCAS operates. Keep 
your fingers away from the claw locks when 
attaching the Upper Part or lifting the patient.
Caution - IV access
Make sure that IV access is not obstructed. 
Caution - do not block the vent holes
Do not cause a blockage of the vent holes 
under the hood since this can cause the 
device to become too hot. 
Caution - device alarms
If there is any malfunction during operation 
the red Alarm LED will illuminate and a 
buzzer signal will be heard. For 
troubleshooting, see section 8.3.
WARNING - MALFUNCTION
If there are interruptions, or the 
compressions are not sufficient, or 
something unusual occurs during operation: 
Push ON/OFF for 1 second to stop LUCAS 
and remove the device. Start manual chest 
compressions.
Caution - do not lift by the Patient Straps
Do not use the Patient Straps to lift the 
patient. The straps are only to attach the 
patient's arms to LUCAS.
3.9 Service
We recommend a yearly servicing of LUCAS 
to make sure that it operates correctly. Use 
the original shipping box when you send 
LUCAS for servicing. Keep the original 
shipping box with padding for this purpose.
WARNING - DO NOT OPEN
Never open the casing of LUCAS. Do not 
change or modify external or internal parts of 
LUCAS.
Unless specified differently, all servicing and 
repairs must be done by service personnel 
that are approved by JOLIFE AB.
If the above conditions are not followed, this 
can lead to patient/operator injury or death, 
and will void the warranty.
Consult your distributor or JOLIFE AB for 
current information on where to send
LUCAS for maintenance.
4 First use preparations
4.1 Delivered items
LUCAS™2 Chest Compression System is 
supplied in one box with:
• A LUCAS device (Upper Part and Back
Plate)
• 3 disposable LUCAS Suction Cups
• A LUCAS Carrying Bag
• Instructions for Use in the relevant
language version
• A rechargeable LUCAS Battery
• A LUCAS Stabilization Strap
• LUCAS Patient Straps
Accessories (optional):
• Disposable LUCAS Suction Cups
• External LUCAS Battery Charger
• Extra LUCAS Batteries
• LUCAS Power Supply with Mains cord
• LUCAS 12-24V DC Car Power Cord
For more accessories, please see appendix 
A: LUCAS™2 parts and accessories.