9 Technical specifications
All specifications in this chapter apply to the LUCAS™2 Chest Compression System.
9.1 Patient parameters
9.2 Compression parameters
9.3 Device physical specifications
9.4 Device environmental specifications
Category
Specifications
Patients eligible for treatment:
Adult patients who fit into the device;
• sternum height of 6.7 to 11.9 inches / 170 to 303
mm
• a maximum chest width of 17.7 inches / 449 mm
The use of LUCAS is not restricted by patient weight.
Category
Specifications
Compression depth (nominal patient)
Patients with sternum height over 7.3 inches / 185 mm:
• 2.1 ±0.1 inches / 53 ±2 mm
Smaller patients with sternum height less than 7.3 inches /
185 mm:
• 1.5 to 2.1 inches / 40 to 53 mm
Compression frequency
102 ±2 compressions per minute
Compression duty cycle
50 ± 5%
Compression modes (operator selectable)
• 30:2 (30 compressions followed
by a 3 seconds ventilation pause)
• Continuous compressions
Category
Specifications
Dimensions when assembled (H × W × D)
22.4 x 20.5 x 9.4 inches / 57 Ч 52 Ч 24 cm
Dimensions Carrying Bag with device inside (H × W × D)
25.6 x 13 x 9.8 inches / 65 Ч 33 Ч 25 cm
Weight of the device with the Battery
17.2 lbs / 7.8 kg
Category
Specifications
Operating temperature
+32°F to +104°F / +0°C to +40°C
- 4°F / -20°C for 1 hour after storage at room temperature
Storage temperature
-4°F to +158°F / -20°C to +70°C
Relative humidity
5% to 98%, non-condensing
IP classification (IEC60529)
IP 43
Operating input voltage
12-24 V DC