8.3 Malfunction alarms
Below is a list of all alarms that can occur on LUCAS. You mute all alarms for 60 seconds if you
push MUTE.
Start with manual compressions immediately if LUCAS does not operate properly.
If the malfunction described above seems permanent, LUCAS must be examined by approved
service personnel. Please consult your local Physio-Control representative or JOLIFE AB.
Contact information is available at www.JOLIFE.com
Reason
Visual LED indication
Audible alarms
Result
Compression pattern outside
limit (too deep, too shallow and
timing failure)
Red alarm LED
Alarm
Compressions stop
Rising temperature in LUCAS
Warning alarm
None
Too high temperature in LUCAS
Red alarm LED
Alarm
Compressions stop
Hardware error
Red alarm LED
Alarm
Compressions stop
Too high Battery temperature
Intermittent red Battery
warning:
The red Battery indication
LED farthest to the right
illuminates intermittently.
Intermittent alarm
Compressions stop
Battery charge too low
Intermittent red Battery
warning:
The red Battery indication
LED farthest to the right
illuminates intermittently.
Intermittent alarm
Compressions stop.
The Battery must be
recharged in the
external Battery
Charger.
Battery error
Constant red Battery
warning:
The red Battery indication
LED farthest to the right
shows a constant light.
Alarm
Compressions stop.
The Battery cannot be
used anymore.