Pulse check – Physio-Control LIFEPAK 1000 User Manual
Page 67
cprMAX™ Technology
LIFEPAK 1000 Defibrillator Operating Instructions
C-3
©2006-2012 Physio-Control, Inc.
Pulse Check
The
PULSE CHECK
option inserts prompting to check for a pulse or check the patient,
depending on the
PULSE PROMPT
setting. The choices for
PULSE CHECK
are:
ALWAYS,
AFTER EVERY NSA, AFTER SECOND NSA
, and
NEVER
. The default setting is
NEVER
.
• The
ALWAYS
option prompts for a pulse check after
CPR TIME 1
and
2
, after a
NO SHOCK
ADVISED
decision, after a single
SHOCK ADVISED
decision with
STACKED SHOCKS OFF
, or
after three consecutive
SHOCK ADVISED
decisions if
STACKED SHOCKS
is
ON
.
• The
AFTER EVERY NSA
option prompts for a pulse check after every
NO SHOCK ADVISED
decision.
• The
AFTER SECOND NSA
option prompts for a pulse check after the second analysis if the
second analysis results in a
NO SHOCK ADVISED
decision, regardless of the first analysis
decision (
SHOCK ADVISED
or
NO SHOCK ADVISED
).
• The
NEVER
option eliminates all
PULSE CHECK
prompts.