Waveshape and measured parameters – Physio-Control LIFEPAK 1000 User Manual
Page 54
Specifications
A-4
LIFEPAK 1000 Defibrillator Operating Instructions
Waveshape and Measured Parameters:
Patient
Impedance (
)
I1 (A)
I2 (A)
I3 (A)
I4 (A)
T1 (ms)
T2 (ms)
25
19.4
10.2
10.1
6.6
7.6
5.1
50
13.2
7.4
7.3
5.0
8.1
5.4
75
10.1
5.8
5.7
4.0
8.3
5.6
100
8.3
4.8
4.8
3.3
8.6
5.7
125
7.0
4.2
4.1
2.9
8.8
5.9
150
6.2
3.7
3.7
2.6
8.8
5.9
175
5.5
3.3
3.3
2.3
8.9
6.0
Note:
Table values are nominal for a 50-joule shock (200 ÷ 4).
T1
T2
I1
I2
I3
I4