Specifications, Defibrillator, Rated energy output – Physio-Control LIFEPAK 1000 User Manual
Page 51

Specifications
LIFEPAK 1000 Defibrillator Operating Instructions
A-1
©2006-2012 Physio-Control, Inc.
SPECIFICATIONS
All specifications are at 20°C (68°F) unless otherwise stated.
Defibrillator
Waveform
Biphasic Truncated Exponential with voltage and duration
compensation for patient impedance.
With Adult Pads:
Patient Impedance Range: 10 – 300 ohms
The following specifications apply from 25 to 175 ohms.
Energy Accuracy:
10% of the energy setting into 50 ohms
15% of the rated energy output into 25 – 175 ohms
Rated energy output is the nominal delivered energy based on the
energy setting and patient impedance, as defined in the following chart.
* Energy setting selected
R ate d E n e rgy O u tpu t
125
150
175
200
225
250
275
300
325
350
375
25
50
75
100
125
150
175
P a tie nt Im pe da nc e
(
ohm s
)
En
e
rg
y
(
J
325 J *
300 J *
275 J *
250 J *
225 J *
200 J *
175 J *
150 J *
360 J *
Rated Energy Output
Energy (J)
Patient Impedance (
)