Defibrillation in manual mode, Analysis, Defibrillation in manual mode -6 – Physio-Control LIFEPAK 1000 User Manual
Page 28: Analysis -6
3-6
LIFEPAK 1000 Defibrillator Operating Instructions
How to Use the LIFEPAK 1000 Defibrillator
DEFIBRILLATION IN MANUAL MODE
The LIFEPAK 1000 defibrillator provides a Manual mode to override the AED features of the
defibrillator. Manual mode provides operator-initiated analysis, charge, shock, and disarm
functions. This mode is useful in a tiered response system when a provider trained in manual
defibrillation and authorized to place the defibrillator in Manual mode takes over the scene from
a BLS-AED trained provider.
To use Manual mode:
1 Press the Menu button.
2 Select
YES
to enter Manual mode. The ECG trace and Heart Rate Indicator appear on
the screen.
3 If the displayed ECG rhythm appears shockable, press
CHARGE
to initiate charging of
the defibrillator. The screen will indicate that the defibrillator is charging and a charge tone
will sound.
4 Clear everyone away from the patient, bed, or any equipment connected to the patient.
5 When the charge is complete, press the flashing
(shock) button to deliver energy to
the patient.
6 After delivering a shock, the energy for each subsequent shock is automatically selected
based on the energy level configured in Setup.
Note: To remove an unwanted charge at any time, press
DISARM
.
Analysis
The LIFEPAK 1000 defibrillator can be set up to display an
ANALYZE
softkey when in
Manual mode.
To initiate an analysis:
1 Confirm that the patient is unresponsive, not breathing, and without a pulse.
2 Press
ANALYZE
.
3 If the rhythm analysis results in a No Shock Advised decision, the defibrillator remains in
Manual mode without further prompts.
4 If the rhythm analysis results in a Shock Advised decision, the defibrillator automatically
begins charging accompanied by a charge tone. If you determine that a shock is not
warranted, press
DISARM
.
5 When the charge is complete, clear everyone away from the patient, bed, or any equipment
connected to the patient.
6 Press the flashing
(shock) button to deliver energy to the patient.
7 After delivering a shock, the defibrillator remains in Manual mode.