Rated energy output, Specifications, Patient impedance ( ) ener gy (j) – Physio-Control LIFEPAK 1000 User Manual
Page 53
Specifications
LIFEPAK 1000 Defibrillator Operating Instructions
A-3
©2006-2012 Physio-Control, Inc.
Waveform
(continued)
With Infant/Child pads:
The following specifications apply from 25 to 175 ohms.
Energy Accuracy (into 50 ohms):
Selected energy
4 +/- 15%; 86 joules +/- 15% maximum
Rated energy output is the nominal delivered energy based on the energy
setting and patient impedance, as defined in the following chart.
* Energy setting selected
R a te d E n e rg y O u tp u t
2 0
3 0
4 0
5 0
6 0
7 0
8 0
9 0
2 5
5 0
7 5
1 0 0
1 2 5
1 5 0
1 7 5
P a tie n t Im p e d a n c e ( o h m s )
En
e
rg
y
(J
3 2 5 J *
3 0 0 J *
2 7 5 J *
2 5 0 J *
2 2 5 J *
2 0 0 J *
1 7 5 J *
1 5 0 J *
3 6 0 J *
Rated Energy Output
Patient Impedance (
)
Ener
gy (J)