Physio-Control LIFEPAK 1000 User Manual
Page 55
Specifications
LIFEPAK 1000 Defibrillator Operating Instructions
A-5
©2006-2012 Physio-Control, Inc.
Electrical Protection:
Input protected against high voltage defibrillator pulses per
IEC 60601-1. See
Figure A-1
Defibrillation–protected, Type BF Patient Connection
Safety Classification:
Internally powered equipment. IEC 60601-1
AED Mode
Shock Advisory
System (SAS):
ECG Analysis system that advises whether a shock is appropriate,
meets rhythm recognition criteria specified in DF80 and IEC
60601-2-4. In AED mode, the device allows a shock only if SAS
advises defibrillation.
Shock Ready Time
Time to first shock (electrodes connected to a patient at power on
and initial rhythm finding is Shock Advised):
• Less than 25 sec to 200 joules
• Less than 30 sec to 360 joules
Energy Sequence:
Multiple levels, configurable from 150 to 360 joules
Shock-to-Shock
cycle time (200J to
300J):
Less than 25 seconds
Time for a 3-shock
sequence
(200J/300J/360J):
Less than 70 seconds
Manual Mode
Energy Sequence
Delivers energy at levels selected in Setup mode.
Charge Time
Charge time:
• 200 joules in less than 7 sec (typical)
• 360 joules in less than 12 sec (typical)
ECG Mode
ECG Display
Provides nondiagnostic ECG display of the patient’s heart rhythm.
Display
Size (Active viewing
area)
120 mm (4.7 in.) x 89 mm (3.5 in.)
Display Type
320 dot x 240 dot LCD with backlight
Frequency Response
0.55 Hz to 21 Hz (-3 dB), nominal
Waveform Sweep
Speed
25 mm/sec for ECG, nominal
Waveform viewing
time
Minimum 4 seconds