Waveshape and measured parameters – Physio-Control LIFEPAK 1000 User Manual
Page 52
Specifications
A-2
LIFEPAK 1000 Defibrillator Operating Instructions
Waveshape and Measured Parameters:
Patient
Impedance (
) I1 (A)
I2 (A)
I3 (A)
I4 (A)
T1 (ms)
T2 (ms)
25
50.3
20.1
19.7
10.7
5.9
3.9
50
28.2
14.6
14.5
9.3
7.5
5.0
75
19.8
11.7
11.7
8.2
8.7
5.8
100
15.5
10.0
9.9
7.3
9.7
6.5
125
12.9
8.7
8.7
6.6
10.4
7.0
150
11.1
7.8
7.7
6.2
11.1
7.4
175
9.8
7.1
7.1
5.7
11.7
7.8
Note:
Table values are nominal for a 200-joule shock.
T1
T2
I1
I2
I3
I4