Ion concentration
9.1
General information
Temperature
measurement with ion
selective measurements
The temperature value is documented with all measurements.
You have the following options to measure the temperature:
Automatic temperature measurement with
– the temperature sensor integrated in the electrode (NTC30 or
Pt1000).
– the integrated temperature sensor of another electrode that is
connected and immersed in the sample at the same time.
– an external temperature sensor (NTC30 or Pt1000).
Manual determination and input of the temperature.
The display of the temperature indicates the active temperature mea-
suring mode:
Operation with burettes
Measurements with ISE methods can be automated by controlling
burettes with the meter:
Connect the burette to the RS232 interface of the ProLab 4000
If necessary, interlink the burettes
(see also operating manual of the burette).
Activate the burette control on the ProLab 4000:
Measuring / Burettes settings... / Activate burettes ()
(see section 9.2.3).
Set the address of the burette on the meter and on the burette
(see page 102 and operating manual of the burette).
Note
Incorrect calibration of ion sensitive electrodes will result in incorrect
measured values. Calibrate regularly before measuring.
Temperature
sensor
Resolution of the
temp. display
Mode
yes
0.1 °C
Automatic with tempera-
ture sensor
-
1 °C
Manual
Note
More information on automation with burette control is available on the
Internet under www.si-analytics.com.