ProLab 4000

Ion concentration

121

ba75646e03

12/2012

15

Change the displayed unit of the measured value with [Unit] if
necessary.

16

If required, display the record of the current measurement with
[Record].

17

Start measuring further samples with [Continue].
Repeat steps 2 - 14 for all samples.
or
Terminate the measuring method with [Cancel] and confirm the
safety query with [OK].
Measurement with the selected method is completed.