10 conductivity, 1 general information, 2 measuring the conductivity – Xylem ProLab 4000 User Manual
Page 139: 1 preparatory activities

ProLab 4000
Conductivity
139
ba75646d03
12/2012
10
Conductivity
10.1
General information
You can measure the following variables:
Conductivity
Specific resistance
Salinity
Total dissolved solids (TDS)
The meter provides the following functions:
AutoRange (automatic selection of the measuring range). If a mea-
suring range is exceeded, AutoRange causes the meter to automat-
ically change to the next higher measuring range and back again.
Therefore, the instrument always measures in the measuring range
with the highest possible resolution.
Stability control to check the stability of the measurement signal.
This ensures the reproducibility of the measuring signal. The back-
ground color is red until the measured value is stable. When the
measured value is stable, the background color switches to green.
Temperature measure-
ment
The LF413T(-ID) and LF313T(-ID) conductivity measuring cells have
an integrated temperature sensor.
10.2
Measuring the conductivity
10.2.1 Preparatory activities
Perform the following preparatory activities when you want to measure:
1
Connect a conductivity measuring cell to the meter.
2
Highlight the channel of the conductivity measuring cell in the
measured value display.
3
If necessary, select the measured parameter with
4
Check whether the setting of the Cell constant is suitable for
the connected conductivity measuring cell. If necessary, cor-
rect the setting.