10 reset, 1 resetting the sensor settings – Xylem ProLab 4000 User Manual
Page 65

ProLab 4000
System functions
65
ba75646e03
12/2012
6.10
Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
6.10.1 Resetting the sensor settings
The sensor settings are reset in the sensor menu.
In the measured value display, activate the relevant sensor window
[Tab] and open the menu with [Menu].
pH
The following settings for pH measurements are reset to the default
settings with the Reset pHMeasuring function:
Cond
The following settings for conductivity measurements are reset to the
default settings with the Reset CondCond-Setup function:
Note
The calibration data is reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Calibration interval...
7 d
Slope in
mV/pH
Measured parameter
pH
pH resolution
High
Zero point
0 mV
Slope
-59.16 mV/pH
Temperature, manual
25 °C
Setting
Default settings
Calibration interval...
150 d
Measured parameter
ϰ
Cell constant
0.475 cm
-1
(Calibrated cell constant)
0.475 cm
-1
(Adjustable cell constant...)
Temperature coeffi-
cient
TC nLF
Tref
Tref 25