Xylem ProLab 4000 User Manual
Page 141

ProLab 4000
Conductivity
141
ba75646d03
12/2012
Stability control
The Stability control function checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of
measured values. The background color is red until the measured
value is stable.
Regardless of the setting for Autom. stability control (see page 63) in
the System menu, you can start the Stability control function manually
at any time.
Criteria
The meter checks the stability of the measured value on the basis of
the temperature measurement. For identical measurement conditions,
the following applies:
1
Highlight the channel of the conductivity measuring cell in the
measured value display.
2
If necessary, select the measured parameter with
3
Freeze the measured value with
The display switches to the display of the selected channel as
necessary.
In the status line, [SC] is displayed.
4
With
As long as the measured value does not meet the stability cri-
terion, the background color is red.
As soon as the measured value is stable, the background color
is green. The current measurement data is downloaded to the
interface. The measured values of all channels meeting the cri-
terion for stability control are marked by SC.
Note
You can terminate prematurely the Stability control function with
nated prematurely, the current measurement data is not downloaded
to the interface.
5
Using
The [SC] status display disappears.
Measured param-
eter
Reproducibility
Response time
Temperature
better than 0.02 °C
> 10 seconds