6 measuring with increment procedure (methods), 1 selecting the measuring method, Measuring with increment procedure (methods) 9.6.1 – Xylem ProLab 4000 User Manual
Page 113: Selecting the measuring method

ProLab 4000
Ion concentration
113
ba75646e03
12/2012
9.6
Measuring with increment procedure (methods)
9.6.1
Selecting the measuring method
The following methods are supported:
Standard addition
Double standard addition
Standard subtraction
Sample addition
Standard subtraction
Blank value addition (Standard addition with blank value correction)
1
Perform the preparatory activities (see page 98).
2
Highlight the channel of the electrode in the measured value
display.
3
If necessary, select the ISE display with
4
If necessary, activate the burette control:
Measuring / Burettes settings... / Activate burettes ()
(see section 9.2.3).
5
Thoroughly rinse the electrode with distilled water.
6
Adjust the temperature of the calibration standard solutions.
7
Select a unit in the menu, Measuring / ISE settings / Units.
8
Select a method in the menu, Measuring / ISE methods.
The selected method is started (see page 114 ff).
Note
The methods described below show the procedure and some displays
of the active burette control.
If the burette control is not active, the [Dosing] button is not displayed.
Any solutions are added manually.