Ion concentration
ProLab 4000
100
ba75646e03
12/2012
Criteria
The criteria of the stability control affect the reproducibility of the mea-
sured values.
If possible, always use the highest reproducibility criteria.
If the stability criteria are not met with your test sample and electrode,
you can change to lower reproducibility criteria. The following criteria
can be adjusted:
High: highest
reproducibility
Medium: medium reproducibility
Low:
lowest reproducibility
Note
With increasing reproducibility, the time until a measured value is eval-
uated as stable is longer.
4
With
While the measured value does not meet the stability criterion,
the background color is red.
As soon as a stable measured value is recognized, the current
measurement data is downloaded to the interface. The mea-
sured values of all channels meeting the criterion for stability
control are marked by SC.
Note
You can prematurely terminate the Stability control function with
nated prematurely, the current measurement data is not downloaded
to the interface.
5
Using
The [SC] status display disappears.