background image

Cond 1970i

Operation

49

ba75692e01

01/2007

4.8

Reset

You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.

Measurement

parameters

The following measured parameters (

InI) are reset to the default

condition:

Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.

Configuration

parameters

The following configuration parameters (InI) are reset to the delivery
status:

Measuring mode

Cell constant

0.475 cm

-1

(calibrated)

0.475 cm

-1

(set up)

Temperature compensation

nLF

Reference temperature

25 °C (TREF25)

Temperature coefficient of the lin-
ear temperature compensation

2.000 %/K

TDS factor

1.00

Baud rate

4800

Interval 1
(automatic saving)

OFF

Interval 2
(for data transmission)

OFF