beautypg.com

3 pc/external printer (rs232 interface), 4 remote control, Pc/external printer (rs232 interface) – Xylem Cond 1970i User Manual

Page 45: Remote control, Section 4.6.3, 1 cts 2 rxd 3 ground 4 txd

background image

Cond 1970i

Operation

45

ba75692e01

01/2007

4.6.3

PC/external printer (RS232 interface)

Via the RS 232 interface, you can transmit the data to a PC or an ex-
ternal printer.
Use the AK340/B (PC) or AK325/S (ext. printer) cable to connect the
interface to the devices.
The data output automatically switches to the RS 232 interface.

Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.

Set up the following transmission data at the PC/printer:

Socket assignment

4.6.4

Remote control

The measuring instrument can be remotely controlled from a PC. This
requires the KOM pilot communication kit. It is available as an acces-
sory.
The instrument is then controlled via commands that simulate key-
strokes and request the current display contents.

Note
A more detailed description is provided within the scope of delivery of
the communication kit.

Baud rate

selectable between: 1200, 2400, 4800, 9600

Handshake

RTS/CTS + Xon/Xoff

Parity

none

Data bits

8

Stop bits

1

1 CTS
2 RxD
3 Ground
4 TxD

1

2

3

4

RS 232

REC