Xylem Cond 1970i User Manual
Page 17
Cond 1970i
Operation
17
ba75692e01
01/2007
Measuring the
temperature
For the temperature compensation, it is required to measure the tem-
perature of the test sample.
You have the following possibilities of measuring the temperature:
z The temperature sensor integrated in the sensor measures the tem-
perature automatically (example: TetraCon 325).
z Automatic temperature measurement by the external temperature
sensor (accessory) NTC30 or Pt1000. This method is required
when using conductivity measuring cells without integrated temper-
ature sensor. Connection of the temperature sensor, see
section 1.3 S
.
z You measure and enter the temperature manually.
Note
The instrument automatically recognizes the type of the used tempera-
ture sensor (NTC30 or Pt1000). The temperature sensor is shown on
the display by TP.
If you use a conductivity measuring cell electrode without a tempera-
ture sensor, proceed as follows:
Note
When determining the cell constant without a temperature sensor, also
set the current temperature of the control standard manually using the
<
▲ > <▼> keys.
1
Measure the current temperature of the test sample using a
thermometer.
2
Enter the temperature on the instrument: Set the temperature
value of the test sample using <
▲ > <▼>.