Setting the temperature compensation TC

The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
Tref20 or Tref25. To switch over the reference temperature, see
section 4.7 C

ONFIGURATION

.

You can select one of the following temperature compensation meth-
ods:

z Nonlinear temperature compensation (nLF)

according to EN 27 888

z Linear temperature compensation (Lin)

with selectable coefficients of 0.001 ... 3.000 %

/

K

z No temperature compensation (- - - -)

Adjusted

temperature

compensation

Each time the instrument is switched on, the adjusted cell constant and
temperature compensation are shown on the display for a short time
(see section 4.1 S

WITCHING

ON

THE

MEASURING

INSTRUMENT

). In order to

view the data, switch the measuring instrument off and switch it on
again.

Note

ips

Select the following temperature compensations given in the table ac-
cording to the respective test sample:

Application tips

Sample

Temperature compensation

Display
indicator

Natural water (ground
water, surface water
and drinking water)

nLF
according to DIN 38404
EN 27 888

nLF

Ultrapure water

nLF
according to DIN 38404
EN 27 888

nLF

Other aqueous solu-
tions

Set linear
temperature coefficient
0.001 ... 3.000 %/K

Lin

Salinity
(seawater)

Automatically nLF according
to IOT

Sal, nLF