4 operation, 1 switching on the measuring instrument, Operation – Xylem Cond 1970i User Manual
Page 15: Switching on the measuring instrument, 4operation
Cond 1970i
Operation
15
ba75692e01
01/2007
4
Operation
4.1
Switching on the measuring instrument
Note
The measuring instrument has an energy saving feature to avoid un-
necessary battery depletion. The energy saving feature switches the
measuring instrument off if no key has been pressed for an hour. The
energy saving feature is not active when the AutoStore function is ac-
tive.
The energy saving feature is also not active
z if the power is supplied by the plug-in power supply,
z if the communication cable and a PC with a running communication
program are connected,
z if the recorder cable is connected,
z if the printer cable is connected (for external printers).
1
Connect a conductivity measuring cell to the measuring instru-
ment.
2
Press the
The display test appears briefly on the display.
Subsequently, the selected cell constant and the temperature
compensation that was set up appear for approx. one second
one after the other.
The measuring instrument then automatically switches to the
measuring mode that was last selected.