1 determining the cell constant (calibrating), Determining the cell constant (calibrating), C t 1 1 – Xylem Cond 1970i User Manual
Page 23: E l l c
![background image](/manuals/310892/23/background.png)
Cond 1970i
Operation
23
ba75692e01
01/2007
4.3.1
Determining the cell constant (calibrating)
Note
This method of automatically determining the cell constant by calibra-
tion with the 0.01 mol/l KCL standard solution can only be used for
measuring cells with cell constants in the range 0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1
.
This is how you can determine the cell constant:
1
Press the
2
Press the
3
Press the
appears on the display: CAL appears on the display.
4
The displayed value is the current calibrated cell constant. You
can:
– accept this setting for measuring with
– continue with step 5 and start a new calibration.
5
Immerse the measuring cell in the control standard solution,
0.01 mol/KCl.
REL1