Operation
Cond 1970i
22
ba75692e01
01/2007
Printing the
calibration protocol
The calibration protocol contains the calibration data of the current cal-
ibration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 39).
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.6.3
before calibrating. After a valid calibration, the record is printed.
Sample printout:
Calibration evaluation
After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the dis-
play.
CALIBRATION PROTOCOL
14.04.02 11:37
Device No.: 99990000
Calibration Conductivity
Cal time: 14.04.02 / 11:37
Cal interval: 180d
Cal Std.: 0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const : 0.478 1/cm
Probe : +++
Display
Cell constant [cm
-1
]
0.450 ... 0.500 cm
-1
0.800 ... 1.200 cm
-1
E3
Eliminate the error according to
chapter 6 W
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1